  <dc:title xml:lang="srp">Značaj korelacije analize gustine kosti i primarne stabilnosti kod izbora dizajna dentalnih implantata</dc:title>
  <dc:creator>Mikić,  Mirko, 1987-, 65029129</dc:creator>
  <dc:description xml:lang="srp">UVOD: Oseointegracija predstavlja preduslov uspjeha implantatne terapije. Kvalitet i kvantitet kosti, dizajn implantata i hirurška tehnika ugradnje, su najznačajniji faktori koji utiču na primarnu stabilnost implantata koja predstavlja preduslov oseointegracije.
METODOLOGIJA: U razrešenju postavljenih ciljeva sprovedena je eksperimentalna
studija na materijalu životinjskog porijekla: goveđem femuru i svinjskim rebrima i klinička prospektivna eksperimantalna studija na ljudima, sa nedostatkom jednog zuba u
bočnoj regiji indikovanih za ugradnju implantata. U ovoj studiji su korišćene dvije
vrste imlantata: samourezujući (Bredent-Narrow SKY) i neurezujući (Nobel Replace –
Conical Connection) istih dimenzija.
REZULTATI: Rezultati eksperimentalne studije su pokazali statistički značajnu
povezanost gustine kosti izražene u HU jedinicama i primarne stabilnosti
samourezujućih i neurezujućih dentalnih implantata izražene u ISQ jedinicama u
kostima goveđeg femura i samourezujućih implantata i kostima svinjskog rebra. Nije
bilo statistički značajne povezanosti neurezujućih dentalnih implantata u kostima
svinjskog rebra. Samourezujući i neurezujući implantati nijesu pokazali statistički
značajnost u primarnoj stabilnosti u kostima različitog kvaliteta. U kliničkoj studiji
je dokazano da postoji statistički značajna jaka pozitivna povezanost između gustine
kosti izražene u HU jedinicama izmjerene u softveru CBCT aparata i primarne
stabilnosti dentalnih implantata izraženih u ISQ jedinicama;
ZAKLjUČAK: Analizom gustine koštanog tkiva CBCT sniimka u softveru aparata
izraženom u HU jedinicama može se predvidjeti stepen primarne stabilnosti
samourezujućih i neurezujućih dentalnih implantata u kostima gustine D1, D2 i
samourezujućih dentalnih imlantata u kostima slabijeg kvaliteta gustine D4. Stepen
primarne stabilnosti neurezujućih dentalnih implantata u kostima slabijeg kvaliteta
D4 ne može se predvidjeti. Više vrijednosti HU povezane su sa višim vrednostima ISQ.</dc:description>
  <dc:description xml:lang="eng">INTRODUCTION: Osseointegration stands as a prerequisite for the success of implant therapy. Bone quality and quantity, implant design, and surgical technique represent the most important factors influencing the primary stability of the implant, which is a prerequisite for osseointegration.
METHODOLOGY: In order to solve the set objectives, an experimental study was conducted on a material of animal origin: bovine femur and pig ribs and a clinical prospective experimental study on humans, lacking one tooth in the lateral region indicated for implant placement. Two
types of implants were used in this study: self-tapping (Bredent-Narrow SKY) and non selftapping (Nobel Replace - Conical Connection) of the same dimensions.
RESULTS: Results of the experimental study showed a statistically significant correlation between bone density expressed in HU units and primary stability of self-tapping and non-selftapping dental implants expressed in ISQ units in bovine femur bones and self-tapping implants
and pig rib bones. There was no statistically significant correlation between non-self-tapping
dental implants in pig rib bones. Self-tapping and non-self tapping implants did not show statistical significance in primary stability in bones of different quality. In a clinical study, it was
demonstrated that there is a statistically significant strong positive correlation between bone density expressed in HU units measured in CBCT apparatus software and primary stability of dental
implants expressed in ISQ units;
CONCLUSION: The analysis of bone density from CBCT images in the software of the apparatus expressed in HU units can be used to predict the degree of primary stability of self-tapping
and non-self-tapping dental implants in bones of density D1, D2 and self-tapping dental implants
in bones of lower quality D4. The degree of primary stability of non self-tapping dental implants in bones of lower quality D4 cannot be predicted. Higher HU values are associated with higher ISQ values.</dc:description>
  <dc:description xml:lang="srp"></dc:description>
  <dc:contributor>Vlahović,  Zoran, 1971-, 58711305</dc:contributor>
  <dc:contributor>Bubalo,  Marija, 1973-, 1646951</dc:contributor>
  <dc:contributor>Lazić,  Zoran, 1963-, 5197415</dc:contributor>
  <dc:contributor>Vasović,  Miroslav, 1982-, 24921191</dc:contributor>
  <dc:date>2020</dc:date>
  <dc:date>2020</dc:date>
  <dc:date>2020</dc:date>
  <dc:date>2020</dc:date>
  <dc:date>2020</dc:date>
  <dc:date>2020</dc:date>
  <dc:type xml:lang="eng">baccalaureate Dissertation</dc:type>
  <dc:format>112 listova</dc:format>
  <dc:format>53472627 bytes</dc:format>
  <dc:identifier>o:1316</dc:identifier>
  <dc:identifier>ID=32457225 ; D-3399</dc:identifier>
  <dc:identifier>thesis:7905</dc:identifier>
  <dc:identifier>cobiss:32457225</dc:identifier>
  <dc:identifier>https://phaidrakg.kg.ac.rs/o:1316</dc:identifier>
  <dc:source>Thesis:7905</dc:source>
  <dc:source>Cobiss:32457225</dc:source>
  <dc:language>srp</dc:language>
  <dc:rights>CC BY-NC-ND 2.0 AT</dc:rights>
  <dc:rights>http://creativecommons.org/licenses/by-nc-nd/2.0/at/</dc:rights>
</oai_dc:dc>